Sign in →

Test Code TNIQUAL Troponin-I  Qualitative ( TVC)

Important Note

This test should only be ordered by TVC 1st Care Providers or after consult with pathologist regarding the option for alternative testing. For other Troponin-I testing please see Troponin I - Fairbanks Memorial Hospital

Additional Codes

Cerner

NextGen

Troponin-I  Qualitative ( TVC)

Troponin-I  Qualitative ( TVC)

Methodology

Abbot i-STAT

Patient Preparation

None

Collection Instructions

  • Standard phlebotomy practices.
  • Venous or Arterial blood only, tubes should be filled to capacity.  

Specimen Requirements

Container

Specimen Type

Tube

Lithium Heparin Plasma

Stability

Storage

Temperature

Stability

Room Temperature

18-28°C

< 30 mins after collection

Rejection Criteria

Not Defined

Result Reporting and Reference Values

Results will be reported as:

  • <0.08 ng/mL: Not Elevated
  • >0.08 ng/mL: Elevated

Reflex Testing

None

Limitations

Interferant

Effect

Hemolysis

Decreased detection on cTnI

Clots

Elevated cTnI levels

Heparin ≥ 90 U/mL

Decrease in cTnI level by approximately 20%.

Time

Less than 4-6 hours from onset of symptoms may yield a normal value.

HAMA or other heterophile antibodies

May interfere with detection of cTnI – consider a different method if in question.

Useful For

Cardiac troponin I measurements are used in the diagnosis and treatment of myocardial infarction and as an aid in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality.

Increased Values for cTnI

  • Myocardial infarction
  • Blunt trauma
  • Myocarditis
  • Sepsis
  • Left ventricular hypertrophy
  • Congestive heart failure

Related Tests