Test Code PLT Platelet Count
Additional Codes
Cerner |
NextGen |
Platelet Count |
Platelet Count |
PLT |
|
Alternate Name(s)
PLT
Patient Preparation
None
Collection Instructions
Do not place samples on a mechanical rocker.
Specimen Requirements
Preferred Volume |
12 x 75 mm tube EDTA-2K or EDTA-3K Lavender filled to capacity |
Minimum Volume |
|
Container |
|
Stability |
|
Rejection Criteria |
|
Result Reporting and Reference Values
Reference Range |
|
Critical Ranges |
|
Reflex Testing
- A fluorescent platelet may be performed when:
- Red cell fragments, microcytic RBC's or white cell cytoplasmic fragments may interfere with automated platelet counts.
- If >10 megakaryocytes/100 WBC are seen on Slide examination
- Slide examination of platelets occurs when:
- Instrument parameter flags indicate abnormal cell populations, requiring a technologist to review slides to verify cell identity and abnormal morphology.
- <50,000/ µL
- >700,000/ µL
- Abnormal Heme Smear Path Review will be performed for platelet counts less than 50,000/ µL or greater than 1,000,000/ µL (first time only).
Limitations
- Giant platelets and clumped platelets may falsely elevate the WBC count and falsely decrease the platelet count.
- Megakaryocytes may falsely increase WBC counts on automated hematology analyzers.
Useful For/ Methodology
Sysmex XN, Hydro Dynamic Focusing (DC Detection) or flow cytometry method using a semiconductor laser with a two-dimensional scattergram for fluorescent platelets (PLT-F).