Sign in →

Test Code HTG2 Thyroglobulin, Tumor Marker, Serum

Useful For

Follow-up of patients with differentiated thyroid cancers after thyroidectomy and radioactive iodine ablation

Reporting Name

Thyroglobulin, Tumor Marker

Specimen Type

Serum Red


Ordering Guidance


For accurate analysis of patients who are known to be thyroglobulin antibody positive, order TGMS / Thyroglobulin Mass Spectrometry, Serum.



Specimen Required


Patient Preparation:  For 12 hours before specimen collection, patient should not take multivitamins or dietary supplements (eg, hair, skin, and nail supplements) containing biotin (vitamin B7).

Supplies: Sarstedt Aliquot Tube, 5 mL (T914)

Collection Container/Tube: Red top (serum gel/SST are not acceptable)

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial.


Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time Special Container
Serum Red Refrigerated (preferred) 7 days
  Frozen  30 days
  Ambient  7 days

Reject Due To

Gross hemolysis Reject
Gross lipemia OK
Gross icterus Reject

Clinical Information

Thyroglobulin (Tg) is a thyroid-specific glycoprotein (approximately 660 kDa) that serves as the source for thyroxine (T4) and triiodothyronine (T3) production within the lumen of thyroid follicles. For T4 and T3 release, Tg is reabsorbed into thyrocytes and proteolytically degraded, liberating T4 and T3 for secretion.

 

Small amounts of intact Tg are secreted alongside T4 and T3 and are detectable in the serum of healthy individuals with levels roughly paralleling thyroid size (0.5-1.0 ng/mL Tg per gram thyroid tissue, depending on thyroid-stimulating hormone [TSH] level). In situations of disordered thyroid growth (eg, goiter), increased thyroid activity (eg, Graves disease), or glandular destruction (eg, thyroiditis), larger amounts of Tg may be released into the circulation.

 

Clinically, the main use of serum Tg measurements is in the follow-up of differentiated follicular cell-derived thyroid carcinoma. Because Tg is thyroid-specific, serum Tg concentrations should be undetectable or very low after the thyroid gland is removed during treatment for thyroid cancer.

 

Current clinical guidelines consider a serum Tg concentrations above 1 ng/mL in an athyrotic individual as suspicious of possible residual or recurrent disease. To improve diagnostic accuracy, it is recommended this measurement be initially obtained after TSH stimulation, either following thyroid hormone withdrawal or after injection of recombinant human TSH. Most patients will have a relatively low risk of recurrence and will thereafter only require unstimulated Tg measurement.

 

If unstimulated (on thyroxine) serum Tg measurements are less than 0.1 to 0.2 ng/mL, the risk of disease is below 1%. Patients with higher Tg levels who have no demonstrable remnant of thyroid tissue might require additional testing, such as further stimulated Tg measurements, neck ultrasound, or isotope imaging. A stimulated Tg above 2 ng/mL is considered suspicious.

 

The presence of antithyroglobulin autoantibodies (TgAb), which occur in 15% to 30% of patients with thyroid cancer, could lead to misleading Tg results. In immunometric assays, the presence of TgAb can lead to falsely low results, whereas it might lead to falsely high results in competitive assays.

 

Traditionally, there have been no reliable means to obtain accurate Tg measurements in patients with TgAb. However, recently trypsin digestion of serum proteins, which cuts both antibodies and Tg into predictable fragments, has allowed accurate quantification of Tg in samples with antibody interferences through measurement of Tg by mass spectrometry. See TGMS / Thyroglobulin Mass Spectrometry, Serum for accurate analysis of patients who are known to be TgAb positive. If TgAb status is unknown, see HTGR / Thyroglobulin, Tumor Marker Reflex, Serum. When HTGR is ordered, TgAb testing is performed first. If TgAb is negative (<1.8 IU/mL), Tg is assayed by immunoassay (sensitive down to 0.1 ng/mL). If TgAb is positive, Tg is assayed by mass spectrometry (sensitive down to 0.2 ng/mL).

Reference Values

Thyroglobulin Tumor Marker

Athyrotic: <0.1 ng/mL

Intact thyroid ≤33 ng/mL

 

Thyroglobulin Antibody:

<1.8 IU/mL

Reference values apply to all ages.

Interpretation

Current guidelines recommend measurement of thyroglobulin (Tg) with a sensitive immunoassay (limit of quantification <1.0 ng/mL); for measurements of unstimulated Tg, the detection limit should be in the 0.1 to 0.2 ng/mL range.

 

In all cases, serum thyroglobulin autoantibodies (TgAb) should also be measured, preferably with a method that allows detection of low concentrations of TgAb. If TgAb are detected, the laboratory report should alert the ordering provider to the possibility of falsely low Tg results if using an immunometric assay. If the apparent Tg concentration is below 1.0 ng/mL, the sample should be remeasured by mass spectrometry. This will allow accurate detection of Tg, in the presence of TgAb, down to 0.2 ng/mL (risk of residual/recurrent disease <1%-3%).

 

Samples from patients with Tg concentrations above 1.0 ng/mL might not require Tg measurement by mass spectrometry because current guidelines suggest further workup might be necessary above this threshold. However, the positive predictive value for residual/recurrent disease is modest when Tg is just above this threshold (3%-25%) in athyrotic patients. Above 10 ng/mL, the risk of residual/recurrent disease is at least 25%, with many studies showing 60% to above 90% risks. In selected patients, therefore, it might also be useful to test TgAb positive samples by mass spectrometry, even if the Tg concentration is above 1.0 ng/mL but not above the 10 ng/mL threshold. These considerations are even more relevant in patients with a known thyroid remnant of a few grams, who may always have serum Tg concentrations of 1.0 to 10 ng/mL, owing to remnant Tg secretion, regardless of the presence or absence of residual/recurrent cancer.

 

It has been determined that the presence of antithyroglobulin autoantibodies (TgAb) in serum can lead to underestimation of Tg concentration by immunometric methods. When TgAb are present in samples with detectable Tg, the Tg values may be underestimated by up to 60% in immunoassays. In addition, approximately 20% of specimens containing TgAb, which are negative for Tg by immunoassay, tested positive by liquid chromatography tandem mass spectrometry. Therefore, measuring Tg by mass spectrometry is the preferred method in TgAb positive patients.

 

The decision levels listed below are for thyroid cancer follow up of athyrotic patients and apply to unstimulated and stimulated thyroglobulin measurements. Decision levels are based on best practice guidelines and the literature, which includes Mayo Clinic studies.

 

Decision levels have not been established but are likely to be somewhat higher for thyroid cancer patients who are not completely athyrotic (ie, patient has some remnant normal thyroid tissue); remnant normal thyroid tissue contributes to serum Tg concentrations 0.5 to 1.0 ng/mL per gram of remnant tissue, depending on the thyrotropin (TSH) level.

 

Tg <0.1 ng/mL: Tg levels must be interpreted in the context of TSH levels, serial Tg measurements, and radioiodine ablation status. Tg levels below 0.1 ng/mL in athyrotic individuals on suppressive therapy indicate a minimal risk (<1%-2%) of clinically detectable recurrent papillary/follicular thyroid cancer.

 

Tg ≥0.1 to 2.0 ng/mL: Tg levels must be interpreted in the context of TSH levels, serial Tg measurements, and radioiodine ablation status. Tg levels 0.1 to 2.0 ng/mL in athyrotic individuals on suppressive therapy indicate a low risk of clinically detectable recurrent papillary/follicular thyroid cancer.

 

Tg 2.1 to 9.9 ng/mL: Tg levels must be interpreted in the context of TSH levels, serial Tg measurements, and radioiodine ablation status. Tg levels 2.1 to 9.9 ng/mL in athyrotic individuals on suppressive therapy indicate an increased risk of clinically detectable recurrent papillary/follicular thyroid cancer.

 

Tg ≥10 ng/mL: Tg levels must be interpreted in the context of TSH levels, serial Tg measurements, and radioiodine ablation status. Tg levels 10 ng/mL or above in athyrotic individuals on suppressive therapy indicate a significant risk (>25%) of clinically detectable recurrent papillary/follicular thyroid cancer.

Day(s) Performed

Monday through Saturday

Report Available

1 to 3 days

Specimen Retention Time

6 months

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

84432

86800

LOINC Code Information

Test ID Test Order Name Order LOINC Value
HTG2 Thyroglobulin, Tumor Marker 57780-9

 

Result ID Test Result Name Result LOINC Value
TGAB2 Thyroglobulin Antibody, S 56536-6
HTGN2 Thyroglobulin, Tumor Marker, S 3013-0
HTG2I Thyroglobulin Interpretation 69053-7

Method Name

Immunoenzymatic Assay

Forms

If not ordering electronically, complete, print, and send 1 of the following forms with the specimen:

-General Request (T239)

-Oncology Test Request (T729)

Secondary ID

62800