Test Code ALKP Alkaline Phosphatase
Additional Codes
Cerner |
NextGen |
Alkaline Phosphatase |
Alkaline Phosphatase |
Alk Phos |
|
Useful For
Alkaline phosphatase measurements are used in the diagnosis and treatment of liver, kidney, bone, parathyroid, and intestinal diseases.
Methodology
Colorimetric (reflectance spectrophotometry) Multiple-point rate
Patient Preparation
None
Collection Instructions
Standard phlebotomy practices.
Specimen Requirements
Container |
|
||||||||||||
Stability |
|
||||||||||||
Rejection Criteria |
Gross Hemolysis |
Result Reporting and Reference Values
Reported in units/L.
Age |
Sex |
Low |
High |
Critical Low |
Critical High |
0-6Y |
M |
95 |
380 |
N/A |
N/A |
F |
95 |
380 |
N/A |
N/A |
|
7-9Y |
M |
175 |
420 |
N/A |
N/A |
F |
175 |
420 |
N/A |
N/A |
|
10-13Y |
M |
135 |
530 |
N/A |
N/A |
F |
130 |
560 |
N/A |
N/A |
|
14-15Y |
M |
135 |
530 |
N/A |
N/A |
F |
70 |
230 |
N/A |
N/A |
|
16-19Y |
M |
65 |
260 |
N/A |
N/A |
F |
50 |
136 |
N/A |
N/A |
|
19-150Y |
M |
38 |
126 |
N/A |
N/A |
F |
38 |
126 |
N/A |
N/A |
Reflex Testing
None
Limitations
- Ortho reports the following biases:
- Bilirubin at 20 mg/dL can cause a 18 units/L bias
- Methotrexate at 200 µg/mL can cause a 29.0 units/L bias
- Nitrofurantoin at 40 µg/mL can cause a 24.0 units/L bias
- Ortho reports no significant effect with the following:
- Lipemia up to 800 mg/dL
- The following interpretive data is visible to the practitioner:
- Nitrofurantoin at blood level of 40 ug/ml (which is extremely high) creates a positive bias of 20% at level of 120 Units/L. Methotrexate at blood level 200 ug/ml creates a positive bias of 22% at level of 130 Units/L